5.1.3. Device Management
Device management spans two pages of device status detailed information:
- Battery
- Coils
- Atrial lead
- RV lead
- LV lead
- Leads implanted
- P Wave Amplitude distribution
- R Wave Amplitude distribution
- Alert parameters setting
- Battery
- Battery status covers the following data:
- Last Battery voltage measurement and date of measurement
- Last Max Energy Charge can be either a capacitor reforming or a max energy shock applied:
- date charged
- energy stored at the beginning of charging
- energy stored at the end of charging
- charge time
- Battery curve: where Recommended Replacement Time (RRT)* and End of Service (EOS)* values are reported.
*: corresponds to the previously used Elective Replacement Indicator (ERI) / End of Life (EOL).
- Leads and Coils
- Leads and coils sub sections report the same data available in the device management summary.
- Notes for each display:
- if the impedance measurement is out of the programmed normal range, the value will be highlighted
- ≤200 or >3000 ohms indicate values outside the range 200 – 3000 ohms instead of exact values
- the measurement date is added when available
- Threshold and Output values will be displayed in grey if the last saved value during a clinic visit.
- Atrial Lead
- Atrial lead sub section presents the following last saved data:
- P Waves (mV) performed beat to beat in millivolts
- Value: measured amplitude
- Measured on: date
- Threshold (V and ms): the last saved pacing threshold value in Volts and milliseconds
- Value: measured amplitude
- Measured on: date
- Output (V and ms): programmed atrial channel output in Volts and milliseconds
- Impedance (ohms): the last impedance measurements (performed every six hours)
Graphs plotted by the daily mean value:
- Detection: 6 month trend of P Wave amplitude
- Impedance: 6 month trend of atrial lead impedance
- RV Lead
- RV lead sub section presents the following last saved data:
- R Waves (mV) performed beat to beat in millivolts
- Value: measured amplitude
- Measured on: date
- Threshold (V and ms): the last saved pacing threshold value in Volts and milliseconds
- Value: measured amplitude
- Measured on: date
- Output (V and ms): programmed RV channel output in Volts and milliseconds
- Impedance (ohms): the last impedance measurements (performed every six hours)
Graphs plotted by the daily mean value:
- Detection: 6 month trend of R Wave amplitude
- Impedance: 6 month trend of RV lead impedance
- LV Lead
- LV lead sub section presents the following last saved data:
- Threshold (V and ms): the last saved pacing threshold value in Volts and milliseconds
- Value: measured amplitude
- Measured on: date
- Output (V and ms): programmed LV channel output in Volts and milliseconds
- Impedance (ohms): the last impedance measurements (performed every six hours)
Graph plotted by the daily mean value:
- Impedance: 6 month trend of LV lead impedance
- Coil
- Coils sub section presents the following last saved data:
- RV Continuity (ohms): coil impedance
- Value: measured (automatically once a week for PARADYM RF/INTENSIA devices and once a day for PLATINIUM devices.)
- Measured on: date
- SVC Continuity (ohms): coil impedance
- Value: measured (automatically once a week for PARADYM RF/INTENSIA devices and once a day for PLATINIUM devices.)
- Measured on: date
- LAST SHOCK
- Impedance (ohms)
- Charge Time (s) in seconds
- Energy (J): energy programmed and delivered in joules
Graph plotted by the daily mean value:
- coil continuity trend: 6 month trend of coils according to the following ranges:
- ≤200 ohms: abnormal
- between 200 ohms and 1000 ohms: normal
- between 1000 ohms and 2000 ohms: elevated
- between 2000 ohms and 3000 ohms: abnormal
- over 3000 ohms: open
- Leads Implanted
- Provides the list of implanted leads and coils like they are entered on the programmer's patient screen.
The following information is captured for each lead:
- Manufacturer
- Model
- Serial Number
- Implant on: date
- P & R Wave Amplitude Distribution
- This section consists of 5 histograms reporting the distribution of atrial and ventricular wave amplitude measured beat to beat:
- Graphs report the total number of cycles counted and the programmed sensitivity (red bar) for the corresponding channel:
- Sinus P waves: recorded during the last 2 months
- A detection in Mode Switch: recorded during the last 2 months
- R waves: recorded since previous in clinic visit
- R waves during VT: recorded since previous in clinic visit
- R waves during VF: recorded since previous in clinic visit
- Alert Parameters
- Shows the list of all alert conditions that can trigger alert transmissions according to their alert activation status the programmed values for programmable parameters.
-
- Alert conditions can be turned ON or OFF in order to trigger Alert transmissions
- Parameters are set in the programmer during a clinic follow up
- Regardless of the ON/OFF setting, detected alerts will be displayed in the Warnings section of the report.